Proline Promass F 300 EtherNet/IP Technical data
223
16.3 Input
Measured variable Direct measured variables
Mass flow
Density
Temperature
Calculated measured variables
Volume flow
Corrected volume flow
Reference density
Measuring range Measuring ranges for liquids
DN Measuring range full scale values 
min(F)
to 
max(F)
[mm] [in] [kg/h] [lb/min]
8 ³⁄₈ 0 to 2 000 0 to 73.50
15 ½ 0 to 6 500 0 to 238.9
25 1 0 to 18 000 0 to 661.5
40 0 to 45 000 0 to 1 654
50 2 0 to 70 000 0 to 2 573
80 3 0 to 180 000 0 to 6 615
100 4 0 to 350 000 0 to 12 860
150 6 0 to 800 000 0 to 29 400
250 10 0 to 2 200 000 0 to 80 850
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the
formula below:
max(G)
= 
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
< 
max(F)
max(G)
can never be greater than 
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
x Constant dependent on nominal diameter
DN x
[mm] [in] [kg/m
3
]
8 ³⁄₈ 60
15 ½ 80
25 1 90