Proline Promass I 100 Modbus RS485 Installation
Endress+Hauser 21
21 mm/m (¼ in/ft) ~2%
1
2
3
4
1
A0016585
6
1 Eccentric clamp connection
2 Line on the underside indicates the lowest point of the eccentric process connection.
3 "This side up" label indicates which side is up
4 Slope the device in accordance with the hygiene guidelines. Slope: approx. 2 % or 21mm/m (0.24 in/feet)
Securing with mounting clamp in the case of hygiene connections
It is not necessary to provide additional support for the sensor for operational performance
purposes. If, however, additional support is required for installation purposes, the
following dimensions must be observed.
Use mounting clamp with lining between clamp and measuring instrument.
A0016588
SI units
DN [mm] 8 15 15 FB 25 25 FB 40 40 FB 50 50 FB 80
A [mm] 373 409 539 539 668 668 780 780 1 152 1 152
B [mm] 20 20 30 30 28 28 35 35 57 57
C [mm] 40 40 44.5 44.5 60 60 80 80 90 90
US units
DN [in] 8 15 15 FB 25 25 FB 40 40 FB 50 50 FB 80
A [in] 14.69 16.1 21.22 21.22 26.3 26.3 30.71 30.71 45.35 45.35
B [in] 0.79 0.79 1.18 1.18 1.1 1.1 1.38 1.38 2.24 2.24
C [in] 1.57 1.57 1.75 1.75 2.36 2.36 3.15 3.15 3.54 3.54