Diagnostics and troubleshooting Proline Promass I 100 Modbus RS485
68 Endress+Hauser
12.9 Event logbook
12.9.1 Event history
A chronological overview of the event messages that have occurred is provided in the
events list which contains a maximum of 20 message entries. This list can be displayed via
FieldCare if necessary.
Navigation path
Event list: F Tool box Additional functions
For information on the event list, see the FieldCare user interface ( 38)
This event history includes entries for:
• Diagnostic events ( 65)
• Information events ( 68)
In addition to the operation time of its occurrence and possible troubleshooting measures,
each event is also assigned a symbol that indicates whether the event has occurred or is
ended:
• Diagnostics event
–
: Event has occurred
– : Event has ended
• Information event
: Event has occurred
To call up the measures to rectify a diagnostic event:
Via "FieldCare" operating tool ( 63)
For filtering the displayed event messages ( 68)
12.9.2 Filtering the event logbook
Using the Filter options parameter, you can define which category of event messages is
displayed in the Events list submenu.
Navigation path
"Diagnostics" menu Event logbook Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)
12.9.3 Overview of information events
Unlike a diagnostic event, an information event is displayed in the event logbook only and
not in the diagnostic list.
Information event Event text
I1000 -------- (device ok)
I1089 Power on
I1090 Configuration reset
I1091 Configuration changed
I1110 Write protection switch changed
I1111 Density adjust. error