11 Operation via FOUNDATION Fieldbus Proline Prowirl 72 FOUNDATION Fieldbus
106 Endress + Hauser
Totalizers -
Failsafe Mode
AUTO - OOS Use this function to define the response of the totalizers to an alarm
condition.
Options:
Stop
The totalizer does not continue to count the flow if a fault is present. The
totalizer stops at the last value before the alarm condition occurred.
Actual Value
The totalizer continues to count the flow on the basis of the current flow
data. The fault is ignored.
Hold Value
The totalizer continues to count the flow on the basis of the last valid flow
data (before the fault occurred).
Factory setting:
Stop
Transducer Block "Totalizer" (Endress+Hauser parameters) / base index 600
Parameter Write access for
operating mode
(MODE_BLK)
Description