Description of device functions Proline Prowirl 73
154 Endress+Hauser
MOLE-% N2 Prerequisite
The function is only available if NATURAL GAS AGA NX-19 was selected in the
NATURAL GAS EQUATION function (ä 149).
Description
Use this function to enter the mole-% of nitrogen in the expected natural gas mixture.
• The values entered in the SPECIFIC DENSITY (ä 153), MOLE-% N2 (ä 154)
and MOLE-%-CO2 (ä 154) functions are interdependent. For this reason, if the
value in one of these functions is changed, the values in the other functions also have
to be adapted accordingly.
• According to the NX-19 standard, the maximum mole-% of nitrogen is 15%.
• The NX-19 standard is not defined for certain combinations of parameters (specific
density, pressure, temperature, mole-% nitrogen and mole-% carbon dioxide) and the
measuring device displays error message #412. In such instances, the mass flow can
no longer be calculated with the NX-19 standard.
User input
5-digit floating-point number
Factory setting
0.0000%
MOLE-%-CO2 Prerequisite
The function is only available if the following was selected in the NATURAL GAS
EQUATION function (ä 149):
• NATURAL GAS AGA NX-19
• NATURAL GAS AGA8 Gross Method 1
• NATURAL GAS SGERG-88
Description
Use this function to enter the mole-% of carbon dioxide in the expected natural gas
mixture.
• The values entered in the SPECIFIC DENSITY (ä 153), MOLE-% N2 (only
NATURAL GAS AGA NX-19) (ä 154), MOLE-% H2 (only NATURAL GAS AGA8
Gross Method 1 and NATURAL GAS SGERG-88) (ä 154) and MOLE-%-CO2
(ä 154) are interdependent. For this reason, if the value in one of these functions
is changed, the values in the other functions also have to be adapted accordingly.
• According to the NX-19 standard, the maximum mole-% of carbon dioxide is 15%.
• The NX-19 standard is not defined for certain combinations of parameters (specific
density, pressure, temperature, mole-% nitrogen and mole-% carbon dioxide) and the
measuring device displays error message #412. In such instances, the mass flow can
no longer be calculated with the NX-19 standard.
User input
5-digit floating-point number
Factory setting
0.0000%
MOLE-% H2 Prerequisite
The function is only available if NATURAL GAS AGA8 Gross Method 1 or NATURAL
GAS SGERG-88 was selected in the NATURAL GAS EQUATION function (ä 149).
Description
Use this function to enter the mole-% of hydrogen in the expected natural gas mixture.
The values entered in the SPECIFIC DENSITY (ä
153), MOLE-% H2 (
ä 154) and
MOLE-%-CO2 (ä 154) functions are interdependent. For this reason, if the value in
one of these functions is changed, the values in the other functions also have to be
adapted accordingly.
User input
5-digit floating-point number
Factory setting
0.0000%
Function description, FLOW COMPUTER group