Proline Prowirl 73 Description of device functions
Endress+Hauser 159
11.17 GAS MIXTURE
Function description, GAS MIXTURE group
Prerequisite
The function is only available if GAS MIXTURE was selected in the SELECT FLUID function (ä 143).
Description
Use this function to program a gas mixture consisting of a maximum of 8 individual gas constituents.
General rules:
At the start, select the number of constituents.
The constituents entered for the gas mixture must have 100.00 Mol-% (=Vol-%) altogether.
A gas and its corresponding Vol-% value can be entered in any order.
If a certain gas is not listed, you can do either of the following:
Program the gas using the real gas equation
Enter the Vol.-% part with the gas type OTHER as well as the following functions:
OPERATING-Z-FACTOR (OTHER) (ä 151), REFERENCE-Z-FACTOR (OTHER) (ä 153) and
REFERENCE DENSITY (OTHER) (ä 152). Thereby the gas calculator with preset gases and the real gas
equation can be combined.
The current gas mixture can be checked in the GAS MIXTURE function.
Programming examples
a. Number of gas constituents: 1
Gas type 1: AIR
Mole-% Gas 1: 100.00%
b. Number of gas constituents: 3
Gas type 1: ARGON
Mole-% Gas 1: 004.00%
Gas type 2: OXYGEN
Mole-% Gas 2: 093.00%
Gas type 3: NITROGEN
Mole-% Gas 3: 003.00%
c. Number of gas constituents: 5
Gas type 1: CARBON DIOXIDE
Mole-% Gas 1: 036.00%
Gas type 2: METHANE
Mole-% Gas 2: 060.00%
Gas type 3: NITROGEN
Mole-% Gas 3: 002.00%
Gas type 4: CARBON MONOXIDE
Mole-% Gas 4: 001.00%
Gas type 5: OTHER
Mole-% Gas 5: 001.00%
OPERATING-Z-FACTOR
1.0000
REFERENCE-Z-FACTOR
1.0000
REFERENCE DENSITY
1.293 kg/m3
NUMBER OF GASES Description
Use this function to enter the number of gases that are used in the gas mixture.
User input
1 to 8
Factory setting
As per the parameter printout supplied with the device if Prowirl 73 was ordered with
a preprogrammed mixture.
Otherwise "1" is the factory setting.