Proline Prowirl R 200 HART Technical data
Endress+Hauser 187
Order code for "Pressure component"
1)
Nominal value
[bar abs.]
Pressure ranges and measured errors
2)
Pressure range
[bar abs.]
Maximum measured error
Option D
Pressure measuring cell 10 bar_a
10 0.01 ≤ p ≤ 2
2 ≤ p ≤ 10
0.5 % of 2 bar abs.
0.5 % o.r.
Option E
Pressure measuring cell 40 bar_a
40 0.01 ≤ p ≤ 8
8 ≤ p ≤ 40
0.5 % of
8 bar abs.
0.5 % o.r.
1) The "mass" sensor version (integrated pressure/temperature measurement) is available only for measuring
devices in HART communication mode.
2) The specific measured errors refer to the position of the measurement in the measuring tube and do not
correspond to the pressure in the pipe connection line upstream or downstream from the measuring
device. No measured error is specified for the measured error for the "pressure" measured variable that can
be assigned to the outputs.
Mass flow saturated steam
Sensor version Mass (integrated temperature
measurement)
Mass (integrated pressure/
temperature measurement)
1)
Process
pressure
[bar abs.]
Flow velocity
[m/s (ft/s)]
Reynolds
number
Window
Measured
value
deviation
PremiumCal
2)
Standard PremiumCal
2)
Standard
> 4.76 20 to 50
(66 to 164)
Re
2
to Re
max
A1 < 1.6 % < 1.7 % < 1.4 % < 1.5 %
> 3.62 10 to 70
(33 to 230)
Re
2
to Re
max
A1 < 1.9 % < 2.0 % < 1.7 % < 1.8 %
In all cases not specified here, the following applies: < 5.7 %
1) Sensor version available only for measuring devices in HART communication mode.
2) Order code for "Calibration flow", option N "0.65% volume PremiumCal 5-point, extended turndown"
Mass flow of superheated steam/gases
4)
Sensor version Mass (integrated pressure/
temperature measurement)
1)
Mass (integrated temperature
measurement) + external pressure
compensation
2)
Process
pressure
[bar abs.]
Flow velocity
[m/s (ft/s)]
Reynolds
number
Window
Measured
value
deviation
PremiumCal
3)
Standard PremiumCal
3)
Standard
< 40 All velocities Re
2
to Re
max
A1 < 1.4 % < 1.5 % < 1.6 % < 1.7 %
< 120 Re
2
to Re
max
A1 < 2.3 % < 2.4 % < 2.5 % < 2.6 %
In all cases not specified here, the following applies: < 6.6 %
1) Sensor version available only for measuring devices in HART communication mode.
2) The use of a Cerabar S is required for the measured errors listed in the following section. The measured error used to calculate the error in the
measured pressure is 0.15 %.
3) Order code for "Calibration flow", option N "0.65% volume PremiumCal 5-point, extended turndown"
4) single gas, gas mixture, air: NEL40; natural gas: ISO 12213-2 contains AGA8-DC92, AGA NX-19, ISO 12213-3 contains SGERG-88 and AGA8
Gross Method 1