Proline Prowirl R 200 HART Installation
Endress+Hauser 27
1
A0019212
1 Maximum insulation height
When insulating, ensure that a sufficiently large area of the housing support remains
exposed.
The uncovered part serves as a radiator and protects the electronics from overheating and
excessive cooling.
NOTICE
Electronics overheating on account of thermal insulation!
Observe the maximum permitted insulation height of the transmitter neck so that the
transmitter head and/or the connection housing of the remote version is completely
free.
Observe information on the permissible temperature ranges.
Note that a certain orientation might be required, depending on the fluid temperature.
6.1.3 Special mounting instructions
Installation for delta heat measurements
Order code for "Sensor version", option CA "mass; 316L; 316L (integrated temperature
measurement), –200 to +400 °C (–328 to +750 °F)"
Order code for "Sensor version", option CB "mass; Alloy C22; 316L (integrated
temperature measurement), –200 to +400 °C (–328 to +750 °F)"
Order code for "Sensor version", option DA "mass steam; 316L; 316L (integrated
pressure/temperature measurement), –200 to +400 °C (–328 to +750 °F)"
Order code for "Sensor version", option DB "mass gas/liquid; 316L; 316L (integrated
pressure/temperature measurement), –40 to +100 °C (–40 to +212 °F)"
The second temperature measurement is taken using a separate temperature sensor. The
measuring device reads in this value via a communication interface.
In the case of saturated steam delta heat measurements, the measuring device must be
installed on the steam side.
In the case of water delta heat measurements, the device can be installed on the cold or
warm side.