Has the correct orientation for the sensor been selected → 22?
• According to sensor type
• According to medium temperature
• According to medium properties (outgassing, with entrained solids)
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping → 22?
Are the measuring point identification and labeling correct (visual inspection)?
Is the device adequately protected against precipitation and direct sunlight?
Are the securing screw and securing clamp tightened securely?
Has the maximum permitted insulation height been observed?
• Has the pressure range been observed→ 192?
• Was the correct orientation selected→ 23?
• Is the pressure unit mounted correctly→ 29?
• Have the pressure gauge valve and the siphon with pressure sensor been mounted using the
prescribed seal and the specified torque→ 29?