10.4.3 Selecting and setting the medium
The Medium selection wizard systematically guides the user through all the parameters
that must be configured in order to select and set the medium.
Navigation
"Setup" menu → Medium selection
Medium selection
Select medium
→  80
Select gas type
→  80
Gas type
→  81
Relative humidity
→  81
Select liquid type
→  81
Steam calculation mode
→  81
Enthalpy calculation
→  82
Density calculation
→  82
Enthalpy type
→  82
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Select medium Select medium type. Gas
Liquid
Steam
Steam
Select gas type The following conditions are
met:
Order code for "Sensor
version",
Option "Mass (integrated
temperature
measurement)"
or
Option "Mass (integrated
pressure/temperature
measurement)"
The Gas option is selected in
the Select medium
parameter parameter.
Select measured gas type. Single gas
Gas mixture
Air
Natural gas
User-specific gas
User-specific gas