Commissioning Proline Prowirl R 200 HART
86 Endress+Hauser
Configuring the pulse output
Navigation
"Setup" menu → Pulse/frequency/switch output
Pulse/frequency/switch output
Assign pulse output 1
→  86
Value per pulse
→  86
Pulse width
→  86
Failure mode
→  87
Invert output signal
→  87
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign pulse output 1 The Pulse option is selected in
the Operating mode
parameter.
Select process variable for
pulse output.
Off
Volume flow
Corrected volume
flow
Mass flow
Total mass flow
*
Energy flow
*
Heat flow
difference
*
Volume flow
Value per pulse In the Operating mode
parameter, the Pulse option is
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→  86):
Volume flow
Corrected volume flow
Mass flow
Total mass flow
*
Energy flow
*
Heat flow difference
*
Enter measured value at which
a pulse is output.
Positive floating-
point number
Depends on country
and nominal
diameter
Pulse width In the Operating mode
parameter, the Pulse option is
selected, and one of the
following options is selected in
the Assign pulse output
parameter (→  86):
Volume flow
Corrected volume flow
Mass flow
Total mass flow
*
Energy flow
*
Heat flow difference
*
Define time width of the
output pulse.
5 to 2 000 ms 100 ms