3.3.6 Installing the wall-mount transmitter housing
There are various ways of installing the wall-mount transmitter housing:
Mounted directly on the wall
Installation in control panel ä 25 (separate mounting set, accessories ä 87)
•Pipe mounting ä 25 (separate mounting set, accessories ä 87)
"
Caution!
Make sure that ambient temperature does not go beyond the permissible range of
– 20 °C to +60 °C (–4 °F to +140 °F), optional – 40 °C to +60 °C (–40 °F to +140 °F).
Install the device in a shaded location. Avoid direct sunlight on the display.
Always install the wall-mount housing in such a way that the cable entries are pointing down.
Mounted directly on the wall
1. Drill the holes as illustrated in the diagram.
2. Remove the cover of the connection compartment (a).
3. Push the two securing screws (b) through the appropriate bores (c) in the housing.
Securing screws (M6): max. Ø 6.5 mm (0.25 inch)
Screw head: max. Ø 10.5 mm (0.4 inch)
4. Secure the transmitter housing to the wall as indicated.
5. Screw the cover of the connection compartment (a) firmly onto the housing.
a0001130-ae
Fig. 18: Mounted directly on the wall
a
b
c
c
90 (3.54)
35 (1.38)
192 (7.56)
81.5(3.2)
mm (inch)