Proline t-mass 65 PROFIBUS DP/PA Installation
Endress+Hauser 25
Installation in control panel
1. Prepare the opening in the panel as illustrated in the diagram.
2. Slide the housing into the opening in the panel from the front.
3. Screw the fasteners onto the wall-mount housing.
4. Screw threaded rods into holders and tighten until
the housing is solidly seated on the panel wall. Afterwards, tighten the locking nuts.
Additional support is not necessary.
a0001131
Fig. 19: Panel installation (wall-mount housing)
Pipe mounting
The assembly should be performed by following the instructions in the diagram.
"
Caution!
If a warm pipe is used for installation, make sure that the housing temperature does not exceed the
max. permitted value of +60 °C (+140 °F).
a0001132
Fig. 20: Pipe mounting (wall-mount housing)
245(9.65)
~110 (~4.33)
210 (8.27)
+0.5 (+0.019)
–0.5(0.019)
+0.5 (+0.019)
–0.5(0.019)
mm (inch)
Ø 2070
(Ø 0.792.75)
~~6.1)155 (
mm (inch)