Installation Proline t-mass 65 PROFIBUS DP/PA
26 Endress+Hauser
3.4 Post-installation check
Perform the following checks after installing the measuring device in the pipe:
Device condition and specifications Notes
Is the device damaged (visual inspection)? –
Does the device correspond to specifications at the measurement point, including
process temperature and pressure, ambient temperature, measuring range, etc.?
Check the name plate.
ä 7
Installation Notes
Correct alignment of pipe/gasket/flowmeter body? ä 13
Professional installation (correct pipe internal diameter, correctly sized gaskets)? ä 13
Is the position chosen for the sensor correct, in other words suitable for sensor
type, fluid properties and fluid temperature?
ä 14
Is there sufficient upstream and downstream pipe sensor? ä 15
Correct installation of flow conditioner (if fitted)? ä 16
Does the arrow on the sensor match the direction of flow through the pipe? ä 14
Correct sensor depth (insertion sensor only)? ä 19
Process environment / process conditions Notes
Is the measuring device protected against moisture and direct sunlight? –
Is the measuring device protected against overheating? ä 18
Is the measuring device protected against excessive vibration? ä 18, ä 110
Check gas conditions (e. g. purity, dryness, cleanliness) Select suitable orientation
ä 14