Proline t-mass 65 Wiring
Endress+Hauser
15
3 Wiring
#
Warning!
Risk of electric shock!
Never mount or wire the measuring device while it is connected to the power supply.
Prior to connecting the power supply, connect the protective earth to the ground terminal on
the housing.
"
Caution!
Risk of damaging the electronic components!
Connect the power supply in accordance with the specifications on the diagram inside the
connection compartments cover.
Additionally for the remote version:
"
Caution!
Risk of damaging the electronic components!
Maximum cable length: 100 m (328 ft)
Observe the cable specifications of the connecting cable Operating Instructions on the
CD-ROM.
!
Note!
Install the connecting cable securely to prevent movement.
Additionally for measuring devices with fieldbus communication:
"
Caution!
Risk of damaging the electronic components!
Observe the cable specification of the fieldbus cable Operating Instructions on the
CD-ROM.
Keep the stripped and twisted lengths of cable shield as short as possible.
Screen and ground the signal lines Operating Instructions on the CD-ROM.
When using in systems without potential matching Operating Instructions on the
CD-ROM.
Additionally for Ex-certified measuring devices:
#
Warning!
When wiring Ex-certified measuring devices, all the safety instructions, wiring diagrams,
technical information etc. of the related Ex documentation must be observed
Ex documentation on the CD-ROM.