3 Wiring
#
Warning!
Risk of electric shock!
• Never mount or wire the measuring device while it is connected to the power supply.
• Prior to connecting the power supply, connect the protective earth to the ground terminal on
the housing.
"
Caution!
Risk of damaging the electronic components!
Connect the power supply in accordance with the specifications on the diagram inside the
connection compartments cover.
Additionally for the remote version:
"
Caution!
Risk of damaging the electronic components!
• Maximum cable length: 100 m (328 ft)
• Observe the cable specifications of the connecting cable → Operating Instructions on the
CD-ROM.
!
Note!
Install the connecting cable securely to prevent movement.
Additionally for measuring devices with fieldbus communication:
"
Caution!
Risk of damaging the electronic components!
• Observe the cable specification of the fieldbus cable → Operating Instructions on the
CD-ROM.
• Keep the stripped and twisted lengths of cable shield as short as possible.
• Screen and ground the signal lines → Operating Instructions on the CD-ROM.
• When using in systems without potential matching → Operating Instructions on the
CD-ROM.
Additionally for Ex-certified measuring devices:
#
Warning!
When wiring Ex-certified measuring devices, all the safety instructions, wiring diagrams,
technical information etc. of the related Ex documentation must be observed
→ Ex documentation on the CD-ROM.