Display functions
Display Function Description
rAtE
Display flow rate Display of the current flow rate or totaliser volume. A negative flow direction is indicated
by a negative digit.
To select the measuring unit press “Set” key
tot
Display totaliser Caution!
Setting of SI/US units is done by miniature switches on the measuring
amplifier board (see page 37)
.
dISP-OF
Display
overflow totaliser
Display of the number of overflows at numerical values >
99 999 999
.
A maximum of 21 overflows is displayed. Starting with the 22nd overflow,
the totaliser begins to sum up from “0” upwards.
rAtE-tot
Display
Flow rate/totaliser
Alternating display (about every 10 seconds) of the current flow rate and totaliser value.
EPd-oFF
Empty Pipe
Detection (EPD)
The EPD function detects whether a measuring pipe is only partially filled with liquid
For ON/OFF switching press “Set” key
EPd-Ad_E
Empty pipe
adjustment
Empty/full pipe adjustment for EPD
To start adjustment press “Set” key
EPd-Ad_F
Full pipe adjustment Note!
Any adjustment has to be done before switching on EPD
(otherwise the
ADJ_ERROR
message is displayed).
During adjustment the message
ADJ_BUSY
is displayed
for about 0.5 s. After any adjustment
ADJ_DONE
is displayed
tESt
Test function Accessing this function activates an automatic test sequence of all display elements;
the following displays are shown:
1.
+ 88 888 888
(incl. display segments)
2.
- 00 000 000
(without display segments)
3. All display elements are blank
4. Flow indication
Additional Information!
Jumper:
Left position auxiliary input configured for “measured value suppression”.
If the measuring value suppression is activated an eight-bar symbol is displayed.
Right position auxiliary input configured for “totaliser reset” to allow resetting the totaliser
to “0” independently of the current display mode. The function of the “totaliser Reset” key is maintained.
In case of system or process errors (incl. EPD) outputs react as described in section 6.1 of the Operating
Manual.
In case of a supply breakdown, all measuring data (e.g. totaliser value) and configurations are saved and
again available once the device is restarted.
If a sensor equipped with an EPD electrode has to be exchanged during servicing, then the EPD calibration
must always be carried out again.