4.3 Connecting the cable of the remote version
Warning!
Danger from electric shock! Switch off the power supply before opening
the instrument.
1. Loosen the safety grip and remove the cover of the
transmitter housing
.
2. Remove the cover from the
connection housing of the sensor
:
– Promag A, H: Loosen all the Phillips screws
– Promag F: Loosen the safety grip and unscrew the cover.
3. Feed both signal and coil-current cable into the appropriate cable entries
of the connection housings.
Caution!
Danger of destroying the coil current control! Only connect or disconnect
the coil cable once the power supply to the instrument has been switched off.
4. Connect the sensor / transmitter cable according to the wiring diagrams
(see Fig. 25).
5. Retighten the connection housing cover securely.
With Promag F, the Allen screw of the safety grip also has to be tightened.