Commissioning Prosonic M FOUNDATION Fieldbus
32 Endress+Hauser
5.8 Envelope curve with device display
After the basic setup, an evaluation of the measurement with the aid of the envelope curve
"envelope curve" (0E) function group) is recommended.
5.8.1 Function "plot settings" (0E1)
Here you can select which information is shown on the display:
envelope curve
substracted signal
mapping
!
Note!
The FAC and the interference echo suppression (map) are explained in BA240F "Prosonic M -
Description of Instrument Functions"
5.8.2 Function "recording curve" (0E2)
This function determines whether the envelope curve is read as
single curve or
cyclic
!
Note!
If the envelope curve mode is active on the display, the measured values are updated in a slower
cycle time. Thus, it is advisable to leave the envelope curve mode after the measuring point has
been optimised.