Prosonic M
Endress+Hauser 17
Installation in narrow shafts
Installation conditions for flow
measurements
Install the Prosonic M at the inflow side, as close above the maximum water level H
max
as possible (take into
account the blocking distance BD).
Position the Prosonic M in the middle of the channel or weir.
Align the sensor membrane parallel to the water surface.
Keep to the installation distance of the channel or weir.
You can enter the "Flow to Level" linearisation curve ("Q/h curve") using FieldCare or manually via the on-
site display.
Example: Khafagi-Venturi flume
L00-FMU4xxxx-17-00-00-en-003
A: Khafagi-Venturi flume, B: inflow, C: outflow, BD: blocking distance, E: empty calibration, F: full calibration,
V: direction of flow
In narrow shafts with strong interference echoes, we
recommend using an ultrasound guide pipe (e.g. PE
or PVC wastewater pipe) with a minimum diameter
of 100 mm (3.94 in).
Make sure that the pipe is not soiled by accumulated
dirt. If necessary, clean the pipe at regular intervals.
L00-FMU4xxxx-17-00-00-en-010
ENDRESS+HAUSER
Prosonic M
ENDRESS+HAUSER
P
rosonic M
ENDRESS+HAUSER
Prosonic M
ENDRESS+HAUSER
Prosonic M
venting
hole
1xb
0
b
0
BD
A
E
BC
V
H
max
=F