EasyManua.ls Logo

Endress+Hauser Prosonic M FMU41 - Commissioning with FOUNDATION Fieldbus configuration tool

Endress+Hauser Prosonic M FMU41
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prosonic M - FOUNDATION Fieldbus Commissioning
Endress+Hauser 55
6.5 Commissioning with a FOUNDATION Fieldbus
configuration tool
!
Note!
For commissioning of the device with a FOUNDATION Fieldbus configuration tool you need to
know the DEVICE_ID, which consists of the following parts:
whereby:
6.5.1 First setup
1. Open the configuration tool and load the Device Descriptions (*.ffo, *.sym and - if required by
the tool - *.cff). Ensure you use the correct files ( ä 38, "Operation with a FOUNDATION
Fieldbus configuration program").
2. The first time it is connected, the device reports as follows:
NI-FBUS-FMU4x-en-001
3. Identify the device using the DEVICE_ID and assign the desired field device tag name
(PD_TAG).
Factory setting: PD_TAG = E+H_PROSONIC_M_XXXXXXXX
Device_ID = 452B481011-XXXXXXXX
452B48 ID code for Endress+Hauser
1011 ID code for Prosonic M
XXXXXXXX Device serial number, as printed on the nameplate