Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning
Endress+Hauser 53
"type" In this linearization mode the measured value is calculated from a linearization table.
The table may consist of up to 32 pairs of values (level - volume). The table must be
monotonically increasing or decreasing.
Selection
•Table
The following additional parameters have to be specified:
Unit of the linearized value, e.g. kg, m3, ft3, ... ("customer unit")
Linearization table ("edit")
"customer unit" Use this parameter to select the desired unit for the linearized values (e.g. kg, m3, ft3,
...). This unit is only indicated on the display. It does not cause a conversion of the
measured value.
If you want to type in a unit, select the option "customer specific".
The parameter "customized text" appears, where you can enter an arbitrary unit
(consisting of up to 5 alphanumeric characters).
"edit" Use this parameter to enter, change or read a linearization table. There are the following
options:
Selection
•Read
The table editor is opened. The existing table can be read but not changed.
•Manual
The table editor is opened. Table values can be entered and changed.
( ä 54, "status table").
•Semi automatic
The table editor is opened. The level is automatically read by the Prosonic S. The
measured value (volume, weight or flow) must be entered by the user.
•Delete
The linearization table is deleted.
Parameter Selection/Remarks
100%
20
mA
0%
0/4
mA
kg
m
ft
3
3