Commissioning
71
6.4.11 "LVL N state" (N = 1 or 2)
"level N" (N = 1 or 2)
Displays the currently measured level.
"act. distance N" (N = 1 or 2)
Dispalys the currently measured distance.
"status"
Use this parameter to define the status of the interference echo suppression.
• enable map
Choose this option in order activate the interference echo suppression. The mapping is then used
for signal evaluation.
• disable map
Choose this option in order to deactivate the interference echo suppression. The mapping is then
no longer used for signal evaluation but it can be reactivated if required.
• delete map
Choose this option in order to delete the mapping. It can not be reactivated again and the
instrument uses the preprogrammed default mapping.
6.5 Envelope curve display
After the basic setup an evaluation of the measurement with the aid of the envelope curve is
recommended, see chapter 10.3.
6.6 After the basic setup
After the basic setup the Prosonic S transmitts the measured value via
• the display module
• the current output
(by default the complete measuring range (0% - 100%) is mapped to the current range [4 - 20mA])
• the HART signal
!
Note!
Sensor management
For instruments with multiple sensor inputs it is possible to deactivate inputs (or sensors) which are
not used. To do so, go to the function "sensor management/FDU sensor N/sensor operation" and
select the desired option:
• on
The sensor is switched on.
• hold
The sensor is switched off. The last measured value is held.
• off
The sensor is switched off. No measured value is transmitted.
Additional parameters are available for optimization of the measuring point. They can be
parametrized as required. A detailed description of all instrument functions is given in the operating
⇒
LVL1 state L100C
level 1:
act. distance 1:
status: