EasyManua.ls Logo

Endress+Hauser Prothermo NMT81 - Page 17

Endress+Hauser Prothermo NMT81
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prothermo NMT81 Installation
Endress+Hauser 17
The following tolerances are applied regardless of an optional WB probe. However, the
position of the flange cannot be adjusted in a welded flange type.
Probe length Tolerance of probe and element positions
1 000 to 25 000 mm (39.37 to 984.25 in) ± 50 mm (1.97 in)
25 001 to 40 000 mm (984.29 to 1 574.80 in) ± 50 mm (1.97 in)
40 001 to 60 000 mm (1 574.84 to 2 362.21 in) ± 100 mm (3.94 in)
60 001 to 100 000 mm (2 362.24 to 3 937.01 in) ± 300 mm (11.81 in)

Other manuals for Endress+Hauser Prothermo NMT81