Prothermo NMT81 Operability
Endress+Hauser 53
a
A0045014
 49 Position of the bottom-point temperature element
a distance between the end of water bottom probe and the 0 mm level point (reference datum plate)
The position of "a" in the figure varies depending on customer's specifications however
this can be configured if necessary.
Liquid temperature
Item Details
Navigation  Operation → Temperature → Liquid temp
Description Shows the average or spot temperature of the measured liquid.
Additional information Read access: Operator
Write access: -
When obtaining data of liquid temperature via NMT81:
Navigate: Setup → Advanced setup → Application → Tank configuration →
Temperature → Liquid temp
Set the source to HART device (NMT81).
NMT element values
Item Details
Navigation Operation → Temperature → NMT element values → Element temperature
Description Displays the NMT element temperature.
Additional information Read access: Operator
Write access: -
Level source
Item Details
Navigation Setup → Advanced setup → Application → Tank configuration → Level → Level
source
Description Defines the source of the level value.
Selection No input value
HART device 1 ... 15 level
Level SR (see Note)