EasyManua.ls Logo

Endress+Hauser RIA14 - 12.6.3 Electrical safety; 12.6.4 Climate class; 12.6.5 Degree of protection; 12.6.6 Vibration resistance

Endress+Hauser RIA14
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIA14 Technical data
Endress+Hauser 29
12.6.3 Electrical safety
As per IEC 61010-1,
UL 61010-1,
CSA C22.2 No. 1010.1-92
12.6.4 Climate class
As per IEC 60654-1, Class C
12.6.5 Degree of protection
IP 66/IP67, Type 4X (not assessed by UL)
12.6.6 Vibration resistance
3g at 2 to 150 Hz as per IEC 60068-2-6
12.6.7 Condensation
Permitted
12.6.8 Installation category
1 as per IEC 61010
12.6.9 Pollution degree
2
12.6.10 Overvoltage category
II
12.6.11 Electromagnetic compatibility (EMC)
CE conformity
Electromagnetic compatibility in accordance with all the relevant requirements of the
IEC/EN 61326 series and NAMUR Recommendation EMC (NE21). For details, refer to the
Declaration of Conformity.
Maximum measured error <1% of measuring range.
Interference immunity as per IEC/EN 61326 series, industrial requirements
Interference emission as per IEC/EN 61326 series, Class B equipment
Connection of the functional grounding may be needed for functional purposes.
Compliance with the electrical codes of individual countries is mandatory.
12.7 Mechanical construction
12.7.1 Design, dimensions
Die-cast aluminum housing for general applications, or optional stainless steel housing

Table of Contents

Other manuals for Endress+Hauser RIA14

Related product manuals