Soliwave FQR56/FDR56
Electrical connection
Endress+Hauser 27
6�3�4 Signal output connection (only FDR56)
A ( 1 : 1 )
B ( 1 : 1 )
2
3
4
5
3
4
5
15 Signaloutputconnection
6�3�4�1 Relay
3 (NO)
4 (CC)
5 (NC)
The following characteristic data apply for the relay signal
output(oatingswitchovercontact):
Switching capacity:
250V/4A, 125 V / 0.4 A or 30 V / 4 A
•Contactmaterial:AgCdO(gold-ashed)
• Switching frequency: max. 4 Hz
Electrical connection Relay
Connection terminals Terminal 3 (NO) Terminal 4 (CC) Terminal 5 (NC)
M12 connector (Binder series 713/763) Connector 2
Terminal 2 (NO) Terminal 3 (CC) Terminal 4 (NC)
Harting connector type HAN8D Connector 2
Terminal 3 (NO) Terminal 4 (CC) Terminal 5 (NC)
•Thecontactmaterialisalsosuitableforswitchingsmall-signalcircuits.However,thisis
possible only if no inductive loads or higher currents have been switched previously.
•Forahighswitchingfrequency,pleaseusethesolid-staterelay.
•WhenusingtheHartingconnectortypeHAN8Dthemaximumswitchingvoltageis120V
or50V.

Other manuals for Endress+Hauser Soliwave FQR56

Related product manuals