Turbimax CUS51D Operation
Endress+Hauser 31
Field of application Range Application Recommended type of
calibration
Titanium dioxide
process sludges
1 to 150 g/l TiO2 [ppm, g/l] Two-point (outside the
process)
Kaolin process water/
process water sludges
0 to 5 g/l Kaolin [mg/l, g/l] Single-point (in the
process)
Preferred applications are highlighted in bold.
Type of calibration (number of calibration points)
One-point calibration
A0050659
29 Single-point calibration
a Factory calibration curve
b New calibration curve
c Calibration point
A single-point calibration causes a change in the slope of the factory calibration curve
programmed into the device.
Two-point calibration
A0050661
30 Two-point calibration
a Factory calibration curve
b New calibration curve
c Calibration points
A two-point calibration causes a change in the slope and zero point of the factory
calibration curve programmed into the device. This type of calibration is recommended as