Waterpilot FMX21 Installation
Endress+Hauser 19
6. Slide the extension cable with the clamping sleeves (item 4) into the adapter (item 5)
7. Fit the cover (item 2) with the sealing ring (item 3) onto the adapter (item 5) and
securely screw together with the adapter.
To remove the cable mounting screw, perform this sequence of steps in reverse.
L
CAUTION
Risk of injury!
‣
Use only in unpressurized vessels.
5.6 Mounting the terminal box
The optional terminal box is mounted using four screws (M4). For the dimensions of the
terminal box, please see the Technical Information TI00431P/00/ EN, "Mechanical
construction" section (see also: www.de.endress.com → Downloads → Media Type:
Documentation).
5.7 Mounting the TMT182 temperature head transmitter with ter-
minal box
1
3
2
6
5
4
3
CLOSE
90°
OPEN
90°
Warning:
Avoid electrostatic charge in explosive atmosphere.
See instructions
Terminal Box for FMX21
1
2
3
4
5
A0018813
1 Mounting screws
2 Mounting springs
3 TMT182 temperature head transmitter
4 Circlips
5 Terminal box
Only open the terminal box with a screwdriver.