Installation Waterpilot FMX21
20 Endress+Hauser
L
WARNING
Risk of explosion!
The TMT182 is not designed for use in hazardous areas.
5.7.1 Mounting the temperature head transmitter:
1. Guide the mounting screws (item 1) with the mounting springs (item 2) through the
bore of the temperature head transmitter (item 3)
2. Secure the mounting screws with the circlips (item 4). Circlips, mounting screws and
springs are included in the scope of delivery for the temperature head transmitter.
3. Screw the temperature head transmitter into the field housing tightly. (Width of
screwdriver blade max. 6 mm (0.24 in))
NOTICE
Avoid damage to the temperature head transmitter.
Do not overtighten the mounting screw.
>7 (0.28)1 2 3
A0018696
1 Terminal box
2 Terminal strip
3 TMT182 temperature head transmitter
Engineering unit mm (in)
NOTICE
Incorrect connection!
A distance of >7 mm (> 0.28) must be maintained between the terminal strip and the
TMT182 temperature head transmitter.