Waterpilot FMX21 Commissioning
Endress+Hauser 39
Description
1 Select the "Level" measuring mode via the
"Measuring mode" parameter.
Menu path: Setup → Measuring mode
L
WARNING
Changing the measuring mode affects the span
(URV)
This situation can result in product overflow.
‣
If the measuring mode is changed, the setting
for the span (URV) must be checked in the
"Setup" operating menu and readjusted if
necessary.
1000 l
0 l
400 mbar
r = 1
g
cm
3
A0018818
1 See table, steps 6 and 7.
2 See table, steps 8 and 9.
2 Select a pressure unit via the "Press eng. unit"
parameter, here "mbar" for example.
Menu path: Setup → Press. eng. unit
3 Select the "In pressure" level mode via the "Level
selection" parameter.
Menu path: Setup → Extended setup → Level →
Level selection.
4 Select a volume unit via the "Output unit"
parameter, here "l" (liters) for example.
Menu path: Setup → Extended setup → Level →
Output unit
5 Select the "Dry" option via the "Calibration mode"
parameter.
Menu path: Setup → Extended setup → Level →
Calibration mode
A0017662
6 Enter the volume value for the lower calibration
point via the "Empty calib." parameter, here 0 liters
for example.
Menu path: Setup → Extended setup → Level →
Empty calib.
7 Enter the pressure value for the lower calibration
point via the "Empty pressure" parameter, here "0
mbar" for example.
Menu path: Setup → Extended setup → Level →
Empty pressure
8 Enter the volume value for the upper calibration
point via the "Full calib." parameter, here
1 000 l (264 gal) for example.
Menu path: Setup → Extended setup → Level →
Full calib.
9 Enter the pressure value for the upper calibration
point via the "Full pressure" parameter, here
400 mbar (6 psi) for example.
Menu path: Setup → Extended setup → Level →
Full pressure