XA02555P-A Waterpilot FMX21
6 Endress+Hauser
Basic specifications
Position 1, 2 (Approval)
Selected option Description
FMX21 UE UK Ex II 2G Ex ia IIC T4...T6 Gb
Position 4 (Probe Tube)
Selected option Description
FMX21 1 316L, d=22mm/0.87in
2 316L, d=42mm/1.66in, flush mount
5 PPS/Polyolefin>316L, d=29mm/1.15in, application salt-water
Optional specifications
No options specific to hazardous locations are available.
Safety
instructions:
General
• Comply with the installation and safety instructions in the Operating
Instructions.
• Staff must meet the following conditions for mounting, electrical
installation, commissioning and maintenance of the device:
• Be suitably qualified for their role and the tasks they perform
• Be trained in explosion protection
• Be familiar with national regulations
• Install the device according to the manufacturer's instructions and
national regulations.
• Avoid electrostatic charging:
• Of plastic surfaces (e.g. enclosure, sensor element, special
varnishing, attached additional plates, ..)
• Of isolated capacities (e.g. isolated metallic plates)
Safety
instructions:
Special conditions
• To avoid electrostatic charging: Do not rub surfaces with a dry cloth.
• In the event of additional or alternative special varnishing on the
enclosure or other metal parts or for adhesive plates:
• Observe the danger of electrostatic charging and discharge.
• Do not install in the vicinity of processes (≤ 0.5 m) generating
strong electrostatic charges.
• Avoid sparks caused by impact and friction.
• Anchor equipment if necessary/secure against swinging.
• Do not use in media or environments which may generate
electrostatic charges on the plastic surfaces.