Description of device parameters Waterpilot FMX21
116 Endress+Hauser
14.20 Expert → Diagnosis → Simulation
Simulation mode
Write permission Operators/Service engineers/Expert
Description Switch on simulation and select the simulation mode. When changing the measuring
mode or the "Lin. mode" level type or when the device is restarted, any simulation that may
be running is switched off.
Options • None
• Pressure →see this table, "Sim. pressure" parameter
• Level, → see this table, "Sim. level" parameter
• Tank content, → see this table, "Sim. tank cont." parameter
• Current, → see this table, "Sim. current" parameter
• Alarm/warning, → see this table, "Sim. error no."
Factory setting None
Transducer block
→
Meas. →
↑ ↓
Sensor → Sensor trim → Position adjustment → Damping → P Level →
PV
→ Current output
↑
↑ ↑
Simulation value
Meas.
• Simulation value level
• Simulation value tank content
Sim.
current
Sim. pressure
Write permission Operators/Service engineers/Expert
Description Enter the simulation value. See also "Simulation mode".
Prerequisite "Simulation mode" = Pressure
Value at switch-on Current pressure measured value
Sim. level
Write permission Operators/Service engineers/Expert
Description Enter the simulation value. See also "Simulation mode".