Waterpilot FMX21 Installation
Endress+Hauser 21
A0018696
Engineering unit mm (in)
NOTICE
Incorrect connection!
‣
A distance of >7 mm (> 0.28) must be maintained between the terminal strip and the
TMT182 temperature head transmitter.
5.8 Mounting the terminal strip for the Pt100 passive
(without TMT182)
If the FMX21 with optional Pt100 is supplied without the optional TMT182 temperature
head transmitter, a terminal strip is provided with the terminal box for the purpose of
wiring the Pt100.
L
WARNING
Risk of explosion!
‣
The Pt100, as well as the terminal strip, are not designed for use in hazardous areas.
A0018815
5.9 Post-installation check
Is the device undamaged (visual inspection)?
Does the device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure
• Ambient temperature
• Measuring range
Are the measuring point identification and labeling correct (visual inspection)?