Diagnostics and troubleshooting Waterpilot FMX21
66 Endress+Hauser
Diagnostic event Reason Corrective measure
Code Description
F283 Memory content Main electronics defective.
Electromagnetic effects are greater than
specifications in the technical data.
The supply voltage is disconnected when writing.
An error occurred when writing.
Perform a reset
F411 Up-/download File is defective.
During the download, the data are not correctly
transmitted to the processor, e.g. because of open
cable connections, spikes (ripple) on the supply
voltage or electromagnetic effects.
Repeat download
Use other file
Perform a reset
F510 Linearization The linearization table is being edited. Conclude entries
Select "linear"
F511 Linearization The linearization table consists of less than 2 points. Table too small
Correct table
Activate table
F512 Linearization The linearization table is not monotonic increasing or
decreasing.
Table not monotonic
Correct table
Activate table
F841 Sensor range Overpressure or low pressure present.
Faulty sensor.
Check the pressure value
Contact Endress+Hauser
Service
F882 Input signal External measured value is not received or displays a
failure status.
Check the bus
Check source device
Check the setting
"M" messages
Diagnostic event Reason Corrective measure
Code Description
M002 Sens. unknown Sensor does not suit the device (electronic sensor
nameplate). Device continues measuring.
Contact Endress+Hauser
Service
M283 Memory
content
Cause as indicated for F283.
Correct measurement can continue as long as you
do not need the peakhold indicator function.
Perform a reset
M431 Adjustment The calibration carried out would result in values that
exceed or fall below the nominal range of the sensor.
Check the measuring
range
Check position
adjustment
Check the setting
M434 Scaling Values for calibration (e.g. lower-range value and
upper-range value) are too close together.
Lower-range value and/or upper-range value exceed
or fall below the range limits of the sensor.
The sensor was replaced and the customer-specific
configuration does not suit the sensor module.
Unsuitable download carried out.
Check the measuring
range
Check the setting
Contact Endress+Hauser
Service
M438 Data set The supply voltage is disconnected when writing.
An error occurred when writing.
Check the setting
Restart the device
M882 Input signal External measured value displays a warning status. Check the bus
Check the source device
Check the setting