What to do if there is no CO gas regulation in Esco Medical Accessories?
- JJimmy YoungAug 18, 2025
If there is no CO gas regulation in your Esco Medical Accessories, check the mains to ensure the system is powered on. If the system is switched off, turn it on. Ensure the CO gas regulator is activated by setting ‘CO ’ to ON in menu 2. Verify the gas supply, ensuring that 0.6 bar (8.70 PSI) of gas pressure is applied and that the correct gas is attached to the CO gas input. Also, check the CO set point to ensure the actual gas concentration is not higher than the set point.