E2608-O2 Rev 27.05.2021
-30...+50 °C
-40…+60 °C
NOTE! Acid gases such as CO
2
and SO
2
will be absorbed by the electrolyte and tend to increase the
flux of oxygen to the electrode. This gives an enhanced oxygen signal of approximately 0.3% of signal
per 1% CO
2
. Electrochemical oxygen sensors are not suitable for continuous operation in
concentrations of CO
2
above 25%.
Other options
Remote probe
Protection IP65, default cable length 3.0 m;
max height 80 mm, max diameter 65 mm