Programming Parameters - Conductivity
The Table below shows the available programming menu items P1 ~ P9. Each parameter is
explained in detail in the subsequent sections.
Parameters
Code Selections
P1 Standard solution series selection
sol
USA (Europe & U.S.A)
CH (China)
P2 Electrode Constant selection
0.1, 1, or 10
P3 Reference Temperature selection
77, 68, and 64
o
F (25,
20,and 18 )
P4
Temperature compensation
coefficient setting
0.00 to 9.99%
P5 Electrode Constant Calibration
P6 Temperature units
℃ ℉
P7 Back light display time setting
0-1-3-6min
P8 Auto power off setting
0-10-20min
P9 Restore to default factory settings OFF / ON
Parameter P1 (Conduc
tivity Solution Setting)
1. From the Conductivity mode, press the MODE key, the ‘P1’ icon appears on the LCD.
2. Use the CAL or the M+/RM keys to toggle through the two (2) selections: USA (for use in the
USA or Europe) and CH (for use in China).
3. Momentarily press MODE to move to the next parameter (P2), or press ENTER to return to the
normal measurement mode.
Parameter P2 (Electrode Constant ‘K’ Selection)
1. From the P2 menu, use the CAL or the M+/RM keys to select the desired electrode constant
(0.1, 1.0, o
r 10). The default setting is K=1.0.
2. Momentarily press MODE to move to the next parameter (P3), or press ENTER to return to the
normal measurement mode.
netzerotools.com
netzerotools.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extended Systems DO700 and is the answer not in the manual?

Extended Systems DO700 Specifications

General IconGeneral
BrandExtended Systems
ModelDO700
CategoryMeasuring Instruments
LanguageEnglish