A-10 ENGLISH
Alarms
ALARM SIGNALS
The F&P 950 Respiratory Humidifier has visual and audible alarms to warn about
interruptions to treatment� These alarms are generated by an intelligent alarm
system, which processes information from the sensors and target settings of the
unit and compares this information to pre-programmed limits�
Alarm type
Mute/unmute alarm
2 minutes�
Text or animation tutorial
Demonstrates solution–
see “alarm conditions”
table on the next page
ALARM CONDITIONS
All possible alarm conditions are listed on the following pages, and all are classified
as medium priority�
As the F&P 950 Respiratory Humidier does not include patient monitoring, these
alarms are considered technical indicators of humidifier performance� It is possible
to have multiple alarm conditions occur simultaneously; under these conditions the
humidifier uses an internal ranking system to display the highest-ranked alarm
All alarms have been designed to be detectable within one meter of the humidifier,
with the alarm signal being three beeps repeated every five seconds
CHECKING ALARM SYSTEM FUNCTIONALITY
To check alarm functionality, remove the heated breathing tube at any time while
the humidifier is powered on but not connected to a patient� This action should
activate the “Disconnection” visual and audible alarms� If either signal is absent, do
not use the humidifier� Contact your servicing department for assistance�
In the event of an unexpected shutdown, the humidifier shall resume the operating
mode and alarm settings (except algorithm-based alarms) prior to the reset if the
interruption is for less than or equal to 30 seconds�