A-10 ENGLISH
Alarms
ALARM SIGNALS
The F&P 950 Respiratory Humidifier has visual and audible alarms to warn about
interruptions to treatment� These alarms are generated by an intelligent alarm
system, which processes information from the sensors and target settings of the
unit and compares this information to pre-programmed limits�
Alarm type
Mute/unmute alarm
2 minutes�
Text or animation tutorial
Demonstrates solution–
see “alarm conditions”
table on the next page�
ALARM CONDITIONS
All possible alarm conditions are listed on the following pages, and all are classified
as medium priority�
As the F&P 950 Respiratory Humidifier does not include patient monitoring, these
alarms are considered technical indicators of humidifier performance� It is possible
to have multiple alarm conditions occur simultaneously; under these conditions the
humidifier uses an internal ranking system to display the highest-ranked alarm�
All alarms have been designed to be detectable within one meter of the humidifier,
with the alarm signal being three beeps repeated every five seconds�
CHECKING ALARM SYSTEM FUNCTIONALITY
To check alarm functionality, remove the heated breathing tube at any time while
the humidifier is powered on but not connected to a patient� This action should
activate the “Disconnection” visual and audible alarms� If either signal is absent, do
not use the humidifier� Contact your servicing department for assistance�
In the event of an unexpected shutdown, the humidifier shall resume the operating
mode and alarm settings (except algorithm-based alarms) prior to the reset if the
interruption is for less than or equal to 30 seconds�