A-12 ENGLISH
Alarms
ALARM CONDITIONS REQUIRED ACTION
The Service Required alarm
activates when the humidifier detects
a potential fault that requires the
humidifier to be serviced
Turn o the humidifier as soon as
appropriate, remove from service, and
contact a technician
Delay: 10 seconds to 5 minutes
The Caution Indicator LED light
illuminates when the humidifier detects
that there is a potential fault with
the humidifier and the screen is not
operational�
Turn o the humidifier as soon as
appropriate, remove from service, and
contact a technician
Delay: <10 seconds
The Cartridge Authentication alarm
activates when the humidifier does not
recognize the sensor cartridge�
If this occurs, the user may choose to
press “Accept” to acknowledge that
the sensor cartridge is not Fisher &
Paykel Healthcare approved� A sensor
cartridge authentication failure icon
will appear at the bottom right of the
display
To remove the sensor cartridge
authentication failure icon, contact
technician to replace sensor cartridge
as soon as appropriate�
The Cartridge Service Life alarm
activates when the humidifier detects
the sensor cartridge has exceeded the
recommended service life�
The sensor cartridge should be
replaced at the next opportunity that
it is safe to do so (when not in use by
a patient)�
Press “Pause Alarm” button to dismiss
the alarm screen
Contact technician to replace sensor
cartridge as soon as appropriate�
Delay: 7 years from the date of
manufacture or after 15,000 hours
of use�
If the alarm is paused, it will reappear
4 hours later�