EasyManua.ls Logo

Fieldpiece HB74 - Functional Specifications - Part 1; Voltage, Resistance, and Capacitance Specs; DC Current Specifications

Fieldpiece HB74
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Functional Specifications
Accuracy specifications good for 75°F±5°F, relative humidity less
than 70%. Accuracy specified as ±% of reading ± number of
least significant digits. For example if the actual parameter is 100
and the accuracy is specified as 1%±3, the measurement could
be as high as 101.3 or as low as 98.7 on the 200 scale.
DC Voltage
Ranges: 200mV, 2000mV, 20V, 200V, 850V
Resolution: 0.1mV on 200mV range
Accuracy: 0.5%±1
Input impedance: 10 M
AC Voltage (50Hz - 500Hz)
Ranges: 200mV, 2000mV, 20V, 200V, 850V
Resolution: 0.1mV on 200mV range
Accuracy: 1.2%±2 @ 50Hz to 500Hz, 2.0%±5 @ 500Hz to
1KHz, 2.0%±5 @ 50Hz to 500Hz on 850VAC range
Input impedance: 10M
Conversion type: Average measuring, rms indicating (sine
wave).
Resistance
Ranges: 200, 2K, 20K, 200K, 20M
Resolution: 0.1 on 200 range
Accuracy: 1%±1 on 200 to 200K ranges, 3%±4 on 20M
range.
Open circuit voltage: <0.3V, <3V on 200 range
Test current: <2mA
Capacitance
Range: 20µF, 200µF, 2000µF, 20KµF.
Accuracy: 4% ±10
Tested at: 3V/21Hz
DC current
Ranges: 20µA, 200µA, 200mA, 20A*
Resolution: 0.01µA
Accuracy: ±(1.0% rdg + 1dgt) on 20µA to 200mA ranges;
±(2.5% rdg + 1dgt) on 20A range
Voltage burden: 250mV on 20µA, 200µA ranges, 400mV on
200mA range and 600mV on 20A range.
* 10A continuous, 20A for 30 seconds maximum.