Operator‘s Manual FERITSCOPE
®
FMP30 Page 109
Measuring
Key sequ. /
Action
Detail of the display Explanation
Press the arrow key to enable the free-
running display mode.
will appear on the display.
[3.00 8.00]: Limits for the analog display
(example).
Place the probe on the specimen and move
it across the surface of the specimen to de-
termine the lateral
-ferrite content distribu-
tion.
[3.00 8.00]: Limits for the analog display
(example).
[ <= =>]: The lower or upper limit of the
analog display is violated
Lift the probe off the specimen.
Use to disable the free-running display
mode.
The last reading taken before the free-run-
ning display ode will be displayed.