Operator‘s Manual FERITSCOPE
®
FMP30 Page 123
Evaluation
Fig. 8-4 Printout of the single readings with tolerance limits enabled (example)
If you change during the display of the block result of the current block to
the block result of a previous block (by pressing ), then the block result of
this block can be printed as well using PRINT. The printout will then include
a list of the readings of this block as well.
8.1.3 Computed parameters - Block result
FERITSCOPE FMP30 Instrument type:
2008-07-25 Current date
Appl. No. Application name/Application number (appears only
if a designation has been assigned ( 5.5 ‘Assigning
Application Designations’, beginning on Page 44)
-- Block result -- Result type
Block No. Current measurement block number/block name (ap-
pears only if a designation has been assigned
5.5 ‘Assigning Application Designations’, begin-
ning on Page 44)
Date Time Date and time of the last measurement of the evaluat-
ed block or closed block, respectively
Mean value Fe. Mean value with a confidence interval of 95 %
Std. dev. s Standard deviation
n in the list of the readings: Sequential number of the
reading
in the block result: Number of the evaluated readings
COV Coefficient of variation
Lowest value Lowest block mean value (Minimum)
Application No. 0 Block No.: 3
n LSL USL
10.2% 10.8%
1| | * | | 10.4
2| | * | | 10.4
3| | * | | 10.4
4| | * | | 10.3
5|<<| | | 10.2