Page 182 Operator‘s Manual FERITSCOPE
®
FMP30
Instrument Settings - Service Function Menu
since this last manual air value capture.
When becomes a static air value measurement necessary?
If the probe contacts the specimen with very different speeds, or if the
touch-down procedure is very slow.
If the probe cannot be lifted off sufficiently from the specimen between
the measurements, e.g., when measurements are made in pipes or
bore holes with inside probes.
Switching on/off the air value acceptance
To keep the measurement errors with static air value measurements as
low as possible, manual air value capture should be carried out at
relatively regular intervals.
Key sequ. /
Action
Detail of the display Explanation
Select Measurement by pressing the arrow
key
or and confirm the selection with
ENTER.
ENTER Select Air value acceptance by pressing
the arrow key
or and confirm the selec-
tion with ENTER.
ENTER
Select the desired acceptance by pressing
the arrow key
or .
The settings of the air value acceptance of
the other Applications will not be affected.
[dynamic]: The air value is captured con-
tinuously during measurement.
[static]: No air value is captured during
measurement. The air value should be cap-
tured manually in regular intervals, see
Page 183.