Page 188 Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Unable to set up,
overwrite or delete
Applications.
Restricted operating
mode is enabled.
Use ON/OFF + EN-
TER to disable the re-
stricted operating mode
with the instrument
switched off before-
hand.
166
Unable to retrieve the
service functions
“Free-running” display
mode is enabled.
Use
to disable the
“free-running” display
mode.
105
Probe does not mea-
sure
Measurement acquisi-
tion did not occur be-
cause the previous
measurement acquisition
was less than two sec-
onds earlier.
Repeat measurement. 34
Computing the coating
thickness is not possible
because the coating is
too thick.
Use a different speci-
men.
--
A wrong probe is con-
nected (Application was
set up with a different
probe).
The indicator for the
measurement method
flashes on the display.
Connect the correct
probe or assign a new
probe.
25
35
Wrong base material
(measurements with
magnetic induction
probes are possible only
on ferromagnetic base
materials).
Use a specimen with a
ferromagnetic base ma-
terial.
--
Malfunction/Message Cause Correction