Page 192 Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Instrument-internal error. Use ON/OFF to turn the
instrument off and then
again on.
If this error occurs re-
peatedly: Inform cus-
tomer service.
28
The internal instrument
memory is full.
Delete the measure-
ment data stored in the
Application or delete a
complete Application.
102
Unable to display read-
ing because it is out of
the probe's measure-
ment range.
Perform measurements
on specimens with
-fer-
rite contents that can be
measured with the con-
nected probe or use a
different probe.
--
Cause: Measurement
carried out incorrectly.
Perform the measure-
ment correctly (e.g., do
not allow the probe to
hover over the speci-
men before or after the
measurement; observe
the minimum lift-off dis-
tance!)
34
Cause: Measurement
with eddy current probe
on a non-ferromagnetic
base material after a
normalization on a fer-
romagnetic base materi-
al.
Perform a normaliza-
tion with the eddy cur-
rent probe on a non-
ferromagnetic base ma-
terial and repeat the
measurement.
61
Malfunction/Message Cause Correction
start
on
Page
E 001
Math Error !
E 004
Appl. memory
overflow !
E 006
Measurements
out of range !