EasyManua.ls Logo

Fisher FERITSCOPE FMP30 - Assigning a New Probe

Fisher FERITSCOPE FMP30
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 30 Operator‘s Manual FERITSCOPE
®
FMP30
System Setup, Maintenance and Cleaning
Detail of the display Explanation of the displays after power up
No reading appears after powering up the instrument be-
cause the last not closed block does not contain any read-
ings.
If the settings in the Service function Storage mode are [do
not save] or [delete upon off], no reading will appear
upon power up as well because the readings have either
not been stored at all or have been deleted when the instru-
ment was switched off.
( 10.8 ‘Storage Mode’, beginning on Page 184)
A designation (in this case “sheet”) has been assigned to
the open Application.
( 5.5 ‘Assigning Application Designations’, beginning
on Page 44)
Once a designation has been assigned to an Application,
it appears in the information lines of the display, where ap-
plicable alternating with the Application number.
The tolerance limits are enabled in the open Applica-
tion:
( 5.6.1 ‘Tolerance Limits’, beginning on Page 46 and
7.3.5 ‘Measurements With Tolerance Limits Enabled’, be-
ginning on Page 91).
The open Application is set to automatic block creation
( 5.6.3 ‘Automatic Block Size and Block Creation’, be-
ginning on Page 49 and 7.3.6 ‘Measurements With a
Fixed Block Size’, beginning on Page 91).
[n=]: Number of the single readings stored in the current /
Block size

Table of Contents