Page 46 Operator‘s Manual FERITSCOPE
®
FMP30
Applications
5.6.1 Tolerance Limits
Having the tolerance limits enabled allows for a simple and quick determi-
nation if the measured -ferrite contents are within specified limits
( 7.3.5 ‘Measurements With Tolerance Limits Enabled’, beginning on
Page 91).
Key sequ. /
Action
Detail of the display Explanation
MENU Use MENU to start Settings.
ENTER Use ENTER to confirm the “tolerance lim-
its”.
or
To activate the tolerance limits:
use the arrow keys to select limits “on” or
to deactivate the tolerance limits: Select lim-
its “off”.
ENTER Use ENTER to confirm the value for the
lower specification limit or:
specify a new lower limit by pressing
or
.
ENTER
or
Use ENTER to confirm the value for the up-
per specification limit or:
specify a new upper limit by pressing
or
.
It is also possible to set the upper and lower specification limits by
making a measurement on an object that corresponds approxi-
mately to the specification limits to be set.
ENTER Press ENTER to confirm the entry.
The tolerance limits are now set and en-
abled. Press DEL to return to the measure-
ment screen.
The instrument is ready to make measure-
ments.