Page 50 Operator‘s Manual FERITSCOPE
®
FMP30
Applications
The block size can be modified even if automatic block creation has al-
ready been enabled. However, the prompt for deleting previously made
measurements will appear4 again because the subsequent statistical eval-
uation does not permit blocks containing different numbers of readings.
5.6.4 Outlier rejection
With outlier rejection enabled, recognized outliers will be indicated on the
display and announced through an audible signal
( 7.3.7 ‘Measurements With Outlier Rejection Enabled’, beginning on
Page 93).
The following methods are available for the outlier rejection function:
Grubbs Test ( Glossary)
Specification of a known spread (Sigma) ( Glossary)
or
Use the arrow keys to set the desired block
size.
ENTER Press ENTER to confirm the selection and
return to the menu to enter additional set-
tings.
Press DEL to return to the measurement
screen.
The instrument is ready to make measure-
ments.
Key sequ. /
Action
Detail of the display Explanation
MENU
Use MENU to start Settings.
Press the arrow key
, until “Outlier Re-
ject” is highlighted on the display.
Key sequ. /
Action
Detail of the display Explanation