Page 52 Operator‘s Manual FERITSCOPE
®
FMP30
Applications
5.6.5 Unit
-ferrite content displayed as -Martensit content. The measured -ferrit
content in Fe% will be computed into the corresponding -Martensit con-
tent. Further information see Chapter Glossary, section -Martensit con-
tent.
5.6.6 Offset
The determined measured value is corrected by the defined offset value.
The offset value can be between -9999 and +9999 μm. The offset value
will always subtracted from the measured value. A positive Offset value de-
creases the displayed value, a negative Offset value increases the dis-
played value.
Key sequ./
Action
Detail of display Explanation
MENU
Use MENU to start Settings.
Press the arrow key
, until “Unit” is high-
lighted on the display.
ENTER Use ENTER to confirm the selection “Unit”.
or
Use ENTER to confirm the unit „Fe%“
or
Press the arrow key
to specify the unit -
Martensit content „
M%“.
ENTER
Press ENTER to confirm the selection/entry
and return to the menu to enter additional
settings.
Press DEL to return to the measurement
screen.
The instrument is ready to make measure-
ments.