Operator‘s Manual FERITSCOPE
®
FMP30 Page 93
Measuring
7.3.7 Measurements With Outlier Rejection Enabled
When making measurements with outlier rejection enabled:
the outlier measurements recognized by the instrument will appear on
the display and will be indicated acoustically ( 5.6.4 ‘Outlier rejec-
tion’, beginning on Page 50),
two long audible signals will sound to indicated the outlier measure-
ment in addition to the audible signal for the measurement acquisition,
recognized outliers will not be integrated in the evaluation of the cur-
rent block or the current Application.,
only the signals for indicating the outlier measurements will sound if the
measurement acquisition signal has been disabled,
outlier measurements recognized in the auto-averaging mode will not
be integrated in the measurement calculation, and
only the audible signals indicating an outlier measurement will sound
when an outlier is recognized (a violation f the tolerance limits will not
be indicated).
Detail of the display Explanation
and will appear in front of the reading on
the display after a measurement has been taken
that has been recognized by the instrument as an
outlier,
Detail of the display Explanation
Outlier! appears briefly in the information lines of the dis-
play in addition to the audible signal if a past reading
with the number n (in the example, n = 1) is recognized
as an outlier measurement (thereafter, the measurement
just taken will appear).